GF-UE190

Type
Ultrasonic Videoscope for EUS
Scope Model
GF-UE190
Product Group
Endoscopic Ultrasound

Channel Length & Diameter

Working Length (mm)
1250
Channel ID (mm)
2.2
Insertion Tube OD (mm)
10.9

Valves [1]

Biopsy Valve, single use
MAJ-1555
Biopsy Valve, reusable
MB-358
Suction Valve, reusable
MAJ-1442

Manual Cleaning Adapter [1]

Suction Cleaning Adapter
MH-856
Channel Plug
MAJ-621
Injection Tube
MH-946
AW Channel Cleaning Adapter
MAJ-629
Water Resistant Cap
MAJ-2295

Manual Cleaner [1]

Manual Cleaner, general indication
Low-foaming, neutral pH detergent which may or may not contain enzymes.
Commercial Product
EndoPreZyme [5]

Olympus Cleaning Brushes [1]

Cleaning Brush, single use [2]
BW-412T, BW-400L
Cleaning Brush, reusable [3]
MH-507, BW-20T

Manual Disinfection [1]

Manual Disinfection, general indication [4]
2% - 3.5% Glutaraldehyde

Olympus EWD [1]

Adapter ETD 3/4
WD00105A (LT), E0424442 (3EUS-40/160)
RFID-Tag ETD 3/4
WD00145A
Adapter ETD Double
WD00122A
Expansion Kit ETD Double
WD00163A
ETD3/4 + Olympus GA
Compatible
ETD3/4 + Olympus PAA
Compatible
ETD Double + Olympus PAA
Compatible
Adapter ETD Basic
WD00312A, WD00325A
Adapter ETD Premium
WD00312A, WD00325A
ETD Basic + Olympus PAA
Compatible
ETD Basic + Olympus GA
Compatible
ETD Premium + Olympus PAA
Compatible
ETD Premium + Olympus GA
Compatible

Olympus OER [1]

OER-AW
MAJ-1508, MAJ-840 (basket)
OER-Smart
MAJ-2111, MAJ-2112, MAJ-2118, MAJ-2127

PROTECH Tip Protector [1]

Size L
Compatible
Size XL
Compatible

Drying Cabinet [1]

EDC
Olympus 4
EDC Plus
Olympus 4
  1. Products listed vary depending on the sales region. Please contact your Olympus representative if you have questions about the availability of Olympus products in your area.

  2. Wherever the cleaning brush MAJ-1888 (single use) is listed, the use of this cleaning brush is obligatory.

  3. Wherever the reusable cleaning brush BW-17K is mentioned, the additional use of the listed single use cleaning brush is obligatory.

  4. The information of the general indication of manual disinfection vary depending on your country. Please check your regional requirements and/or contact your Olympus representative if you have any questions.

  5. Detergent intended prior to automated reprocessing in endoscope washer-disinfector (EWD).

This endoscope is compatible with ETD endoscope washer-disinfector that operates with Olympus GA or Olympus PAA Pro chemicals. Wear and tear of the endoscopes might be shown by cracked, peeled or discolored adhesives at either end of the bending section and around objective lens and light guide lens. In case of using PAA Pro chemicals, the signs of degradation may occur after 200 reprocessing processes or later. In case of such degradation, maintenance of the endoscopes will be required. For more details, contact Olympus.