UM-2R
- Type
- Ultrasonic Miniature Probes
- Scope Model
- UM-2R
- Product Group
- Endoscopic Ultrasound
Channel Length & Diameter
- Working Length (mm)
- 2050
- Insertion Tube OD (mm)
- 2.5
Manual Cleaning Adapter [1]
- Water Resistant Cap
- MH-244
Manual Cleaner [1]
- Manual Cleaner, general indication
- low-foaming, neutral pH detergent or enzymatic detergent
- Commercial Product
- EndoPreZyme [5]
Olympus EWD [1]
- ETD Mini + Olympus GA
- Compatible
- ETD3/4 + Olympus GA
- Compatible
- RFID-Tag ETD Basic
- E0424431
- RFID-Tag ETD Premium
- E0424431
- ETD Basic + Olympus PAA
- Compatible
- ETD Basic + Olympus GA
- Compatible
- ETD Premium + Olympus PAA
- Compatible
- ETD Premium + Olympus GA
- Compatible
Olympus OER [1]
- OER-2
- MAJ-1763
- OER-3
- MAJ-1764
- OER-4
- MAJ-1764
- OER-5
- MAJ-1764
- OER-6
- MAJ-1764
- OER-AW
- MAJ-1764
- OER-Pro
- MAJ-1764
- OER-Elite
- MAJ-2119
- OER-Smart
- MAJ-2119
Products listed vary depending on the sales region. Please contact your Olympus representative if you have questions about the availability of Olympus products in your area.
Wherever the cleaning brush MAJ-1888 (single use) is listed, the use of this cleaning brush is obligatory.
Wherever the reusable cleaning brush BW-17K is mentioned, the additional use of the listed single use cleaning brush is obligatory.
The information of the general indication of manual disinfection vary depending on your country. Please check your regional requirements and/or contact your Olympus representative if you have any questions.
Detergent intended prior to automated reprocessing in endoscope washer-disinfector (EWD).