UM-2R

Type
Ultrasonic Miniature Probes
Scope Model
UM-2R
Product Group
Endoscopic Ultrasound

Channel Length & Diameter

Working Length (mm)
2050
Insertion Tube OD (mm)
2.5

Manual Cleaning Adapter [1]

Water Resistant Cap
MH-244

Manual Cleaner [1]

Manual Cleaner, general indication
low-foaming, neutral pH detergent or enzymatic detergent
Commercial Product
EndoPreZyme [5]

Manual Disinfection [1]

Manual Disinfection, general indication [4]
2% - 3.5% Glutaraldehyde

Olympus EWD [1]

ETD Mini + Olympus GA
Compatible
ETD3/4 + Olympus GA
Compatible
RFID-Tag ETD Basic
E0424431
RFID-Tag ETD Premium
E0424431
ETD Basic + Olympus PAA
Compatible
ETD Basic + Olympus GA
Compatible
ETD Premium + Olympus PAA
Compatible
ETD Premium + Olympus GA
Compatible

Olympus OER [1]

OER-2
MAJ-1763
OER-3
MAJ-1764
OER-4
MAJ-1764
OER-5
MAJ-1764
OER-6
MAJ-1764
OER-AW
MAJ-1764
OER-Pro
MAJ-1764
OER-Elite
MAJ-2119
OER-Smart
MAJ-2119
  1. Products listed vary depending on the sales region. Please contact your Olympus representative if you have questions about the availability of Olympus products in your area.

  2. Wherever the cleaning brush MAJ-1888 (single use) is listed, the use of this cleaning brush is obligatory.

  3. Wherever the reusable cleaning brush BW-17K is mentioned, the additional use of the listed single use cleaning brush is obligatory.

  4. The information of the general indication of manual disinfection vary depending on your country. Please check your regional requirements and/or contact your Olympus representative if you have any questions.

  5. Detergent intended prior to automated reprocessing in endoscope washer-disinfector (EWD).