Type Testing Washer-Disinfectors (AERs/EWDs), Part 3

Type testing washer-disinfectors for reprocessing thermolabile endoscopes (EWDs) in accordance with the revised EN ISO 15883 series of standards

Part 3: Scope and conduct of testing for process chemicals not included by the EWD manufacturer in the type test

This article is Part 3 of a series on type testing washer-disinfectors for thermolabile endoscopes (AERs/EWDs). It focuses on the evaluation and validation of additional process chemicals that were not included in the original AER/EWD type test.

To minimize unnecessary testing while maintaining process safety and performance, the authors propose a tiered validation approach consisting of three levels:

  • laboratory testing without the AER/EWD

  • testing within the AER/EWD

  • testing in the AER/EWD using surrogates and representative endoscopes.

This strategy aims to avoid repeating the complete Annex C type test while still demonstrating the safety and effectiveness of the modified process.

For detergents, the article recommends comparative laboratory cleaning studies against the detergent used during the original type test. These evaluations should employ established test soils, such as modified biofilm and blood-based soils, and be complemented by assessments within the AER/EWD to verify the impact on key process parameters, including flushing pressure. Cleaning performance should then be confirmed using selected surrogates and representative endoscopes from different product families.

For disinfectants, the authors emphasize that in vitro efficacy testing must first demonstrate the required antimicrobial activity under the intended process conditions, including temperature, exposure time, and concentration. In addition, the wettability of the load should be assessed, for example, by comparing surface tension, to ensure adequate surface coverage and effective disinfection.

The article further highlights the importance of verifying that additional process chemicals do not adversely affect other AER/EWD functions or essential performance requirements. These include material compatibility, biocompatibility of chemical residues, residue limits and detection methods, chemical self-disinfection, dosing accuracy, channel flow control, temperature regulation, water quality, and compatibility with both the endoscope and the AER/EWD.

Finally, the authors conclude that if equivalence to the originally validated process cannot be demonstrated, or if uncertainties arise during extrapolation or interpretation of the results, a full AER/EWD type test may be required. They also emphasize that close collaboration between AER/EWD manufacturers, endoscope manufacturers, and process chemical suppliers remains essential throughout the product lifecycle, even after the initial type testing has been completed.

Read the full article to gain deeper insights into the proposed validation strategy, testing methodology, and practical considerations for introducing additional process chemicals following AER/EWD type testing.